Z-0775-2014Class II

Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision...

Lab Vision Corp. / initiated 2013-12-06 / Terminated

Lab Vision Corporation began a recall of Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision Corporation, Fremont, CA Thermo Fisher Scientific... on . FDA classified it as Class II and gives the reason as "Reports that the product is not performing as expected. In house testing showed no staining on breast cancer tissues". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0775-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Lab Vision Corp.
Product
Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision Corporation, Fremont, CA Thermo Fisher Scientific Anatomical Pathology Tudor Rd., Manor Park Rancorn, Cheshire WA7 1TA UK Analyte specific reagent. It is intended for qualitative immunohistochemistry with normal and neoplastic formalin-fixed, paraffin embedded tissue sections, to be viewed by light microscopy.
Product code
MVU Reagents, Specific, Analyte
Reason for recall
Reports that the product is not performing as expected. In house testing showed no staining on breast cancer tissues.
FDA root cause
Under Investigation by firm
Action
Lab Vision Corporation part of Thermo Fisher Scientific sent an Urgent Medical Device Recall letter dated December 6, 2013, to all affected customers. Revised letters will be sent on January 8, 2014. Changes include that the reagent is an Analyte Specific Reagent (ASR) and not an in vitro diagnostic device in the US. They also clarify that a false positive was reported by a customer and a false negative was confirmed during in house testing. Letters also correct the expiry date from 12/2013 to 03/2015. Data sheet was also updated to eliminate information stating that the presence of ER in breast tumors indicates an increased likelihood of response to estrogen therapy and also that the positive control must be determined by the investigator. If you have any questions, please contact Technical Service at 1-800-522-7270, option 2, option 7.
Quantity in commerce
5
Distribution
Worldwide Distribution: US (nationwide) including the states of TX and MA and internationally to: Turkey.
Product codes and lots
PN/ MS-354-P0, LN 354P1206K; Expiration 03/2015.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0775-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0775-2014Class IITerminatedVoluntary: Firm initiated