510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1092-2014Class II

00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter...

Zimmer GmbH / initiated 2014-01-10 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of 00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48... on . FDA classified it as Class II and gives the reason as "This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1092-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm, Sterile, QTY 1, Rx Product Usage: Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.
Product code
KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
Reason for recall
This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Zimmer sent an Urgent Medical Device Recall dated January 10, 2014 to distributors via electronic mail. The hospital/surgeon that has implanted this manufacturing lot will be identified and notified by the distributor. The letter identified the affected product, problem and actions to be taken. Distributors were instructed to locate and remove the affected product in their territory, as well as identifying and notifying the hospital/surgeon that has implanted this manufacturing lot as applicable. The distributor will return on-hand affected products to Zimmer and ensure all of their products are accounted for using the return form provided in the letter. Distributors and hospital consignees are to complete and return the inventory return form and acknowledgement form indicating which hospital/surgeon has implanted the product and to which they have provided notice to. Email a copy of the completed return form to [email protected] outside of the United States will occur approximately two weeks after the United States communications. For questions call 1-877-946-2761
Quantity in commerce
10 units
Distribution
Distribution USA in the state of Illinois.
Product codes and lots
lot 62267250

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1092-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1092-2014Class IITerminatedVoluntary: Firm initiated