Z-0607-2014Class II

IntelliFill i.v. Pharmacy Compounding Device (PCD). Intended to fill syringes from 0.5mL up to 11.5mL using...

Baxter Healthcare Corporation / initiated 2013-11-08 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of IntelliFill i.v. Pharmacy Compounding Device (PCD). Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the... on . FDA classified it as Class II and gives the reason as "Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0607-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
IntelliFill i.v. Pharmacy Compounding Device (PCD). Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the orders.
Product code
NEP System/Device, Pharmacy Compounding
Reason for recall
Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A Service Bulletin was sent to all IntelliFill i. v. customers February 22, 20 I 0 by USPS mail. The Service Bulletin informs customers that IntelliFill i.v. operators are to be trained to listen for the running HEPA filtration fan motor and observe/feel for airflow coming from the automation deck. The field service engineer dispatched to the site replaced the damaged sensor on 11-Feb-2010 and a Service Bulletin was sent to all customers regarding the function and use of the HEPA filter on 22-Feb-2010. A new HEPA sensor was designed and installed into all current IntelliFill i.v. devices between 21-July-2011 to 9-Feb-2012.
Quantity in commerce
28 devices
Distribution
Nationwide Distribution including the states of CA, CO, CT, FL, IL, LA, MA, MD, MI, NC, NE, NY, OH, OR, PA, TN, TX, and WV.
Product codes and lots
Lot # 123, 125R, 127, 129, 145, 146R, 147R, 148, 215R, 500, 501, 502, 503, 504, 505, 506, 508, 510, 511R, 512, 514, 515, 516R, 517, 518, 519, and 520.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0607-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0607-2014Class IITerminatedVoluntary: Firm initiated