Z-0734-2014Class II

Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs...

Remel Inc / initiated 2013-11-21 / Terminated

Remel Inc began a recall of Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 cartridges per pack, 50 discs per... on . FDA classified it as Class II and gives the reason as "Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0734-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remel Inc
Product
Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 cartridges per pack, 50 discs per cartridge. Firm name on the label is Oxoid, Ltd., Basingstoke, United Kingdom. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes.
Product code
JTN Susceptibility Test Discs, Antimicrobial
Reason for recall
Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes.
FDA root cause
Unknown/Undetermined by firm
Action
Oxoid and Remel part of Thermo Fisher Scientific sent an Urgent Medical Device Recall letter dated November 20, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The letter asked customers to destroy the affected product, if any was found on hand. The letter also asked if the customer further distributed the product that they inform their downstream customers. A response form was included with the letter and the letter instructed customers to complete and return the response form to indicate they had received the notification and that they had complied with it. If you have any questions please contact our Technical Service Department at 800-255-6730 (US) or 913-888-0939 (International).
Quantity in commerce
45/5/50 - discs/cartridges/packs
Distribution
Nationwide Distribution including the states of AL, AZ, CA, CO, FL, HI, IA, IL, KY, LA, MD, MI, MN, MO, MS, NC, NY, OH, SC, TN, TX, VA, WI, and WV
Product codes and lots
Lot number: 1223566 EXP: 8/31/2013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0734-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0734-2014Class IITerminatedVoluntary: Firm initiated