Z-0611-2014Class I

Puritan Bennett 840 ventilator systems software part number 4-070212-85, revision AB-AG. The intended use of...

Medtronic, Inc. / initiated 2013-12-16 / Terminated / root cause: software

Covidien began a recall of Puritan Bennett 840 ventilator systems software part number 4-070212-85, revision AB-AG. The intended use of the product is to provide mechanical ventilation... on . FDA classified it as Class I and gives the reason as "Covidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0611-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Puritan Bennett 840 ventilator systems software part number 4-070212-85, revision AB-AG. The intended use of the product is to provide mechanical ventilation to patients.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Covidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the device stops mechanical ventilation due to a software error.
Root cause tag
software
FDA root cause
Software design
Action
Covidien sent an Urgent Medical Device Voluntary Field Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue using their Puritan Bennett 840 ventilators until it is feasible to install the software update discussed in the letter. Covidien has developed a software solution to address this issue. Upon availability, there will be several ways to install the software update. To initiate the process for updating your software and to select the method that best suits your facility, please visit our software update management portal at www.PB840technicalupdate.com and follow the instructions to register your ventilators. If you need further assistance contact our Technical Support Department @ 1-800-255-6774, option 4, then option 1. For questions regarding this recall call 800-255-6774.
Quantity in commerce
16,923 US, 39,372 OUS
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Austria, Estonia, Belgium, Spain, Italy, Poland, Bulgaria, Finland, Portugal, Switzerland, France, Lithuania, Romania, Cyprus, Great Britain, Czech Republic, Greece, Latvia, Slovenia, Germany, Hungary, Slovak Republic, Ireland, Netherlands, Turkey, Croatia, Andorra, Albania, Algeria, Bahrain, Bosnia and Herzegovina, Egypt, Georgia, Iran, Iraq, Israel, Jordan, Kenya, Kuwait, Lebanon, Libya, Macedonia, Montenegro, Morocco, Russia, Saudi Arabia, Serbia, South Africa, Syria, Sudan, Tunisia, UAE, Uganda, Puerto Rico, Mexico, Columbia, Chile, Brazil, Argentina, Uruguay, Panama, Costa Rica, Peru, China, Korea, Japan, Taiwan, Hong Kong, Malaysia, Singapore, Sri Lanka, Australia, New Zealand.
Product codes and lots
Software Part number 4-070212-85, revision AB-AG

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0611-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0611-2014Class ITerminatedVoluntary: Firm initiated