ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems...
Philips Medical Systems (Cleveland) Inc began a recall of ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH on . FDA classified it as Class II and gives the reason as "The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0785-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH
- Product code
- IZJ Camera, X-Ray, Fluorographic, Cine Or Spot
- Reason for recall
- The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On October 30, 2013 & December 4, 2013, the firm sent Urgent Medical Device Recall letters to their customers.
- Quantity in commerce
- 255 units
- Distribution
- Worldwide Distribution: US (nationwide) including states of: AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.
- Product codes and lots
- ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B, 03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348B, M@5807B, M@5874B, M@5894B, M@5932B, M@5941B, M@5963B, M@6041B, M@6062B, M@6229B, M@6354B, M@6477B, M@6485B, M@6546B, M@6555B, M@6692B, M@6754B, M@6809B, M@6815B, M@6948B, M@7007B, M@7023B, M@7093B, M@7194B, M@7198B, M@7207B, M@7214B, M@7280B, M@7285B, M@7286B, M@7408B, M@7540B, M@7596B, M@7675B, M@7704B, M@8156B, M@8324B, M@8344B, M@8429B, M@8523B, M@8578B, M@8588B, M@8637B, M@8639RB, M@8648B, M@8808B, M@8813B, M@8898B, M@8977B, M@9050B, M@9261B, M@9276B, M@9355RB, M@9475B, M@9544B, M@9647B, M@9656B, M@9733B & M@9796B. ADAC ARGUS: Model #88295; Serial #: 9211014, 9301005, 9304019, 9304024, 9304028, 9401028, 9401031, 9401037, 9505014, 9910033, 93-10678- 1292, 9910028P, M@10225AMT, M@3732B, M@3903B, M@3954B, M@3956B, M@3960B, M@4249B, M@4415B, M@4646B, M@4868B, M@5172B, M@5321B, M@7341B & M@AMT1154. ADAC CIRRUS: Model #88294; 34, 718558, 9206009, 9207003, 9207004, 9207005, 9207007, 9211006, 9300113, 9304013, 9309012, 9404022, 9407028, 9509042, 9604098, 9910030, 93-10735-0393, 9621RQ, M@10110AMT, M@10324AMT, M@10402AMT, M@3599G-08, M@3905Q, M@4029Q, M@4129Q, M@4142Q, M@4668Q, M@4743Q, M@5003Q, M@5668Q, M@5895Q, M@AMT126R, M@AMT1282 & M@MVS3227A.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0785-2014 | Class II | Terminated | Voluntary: Firm initiated |