Z-0784-2014Class II

Zimmer Segmental Systems; Polyethylene Inserts, Size C. Knee replacement

Zimmer GmbH / initiated 2013-12-19 / Terminated / root cause: design

Zimmer, Inc. began a recall of Zimmer Segmental Systems; Polyethylene Inserts, Size C. Knee replacement on . FDA classified it as Class II and gives the reason as "After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0784-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Segmental Systems; Polyethylene Inserts, Size C. Knee replacement.
Product code
KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
Reason for recall
After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.
Root cause tag
design
FDA root cause
Device Design
Action
Depth of recall " Notifications were sent to customers, surgeons and risk managers, which use the affected product. Distributors were also informed of the notifications sent to surgeons and risk managers. The customers were identified by performing an invoice search and distribution history for the Polyethylene Inserts (00-5850-012-95 and 00-5850-013-95) since product launch. Recall Communication " All distributors were notified via electronic mail. Distributors that have received affected inventory were also notified via courier. Hospital risk managers and surgeons were also notified via courier. On 12/19/13, hospital risk managers and surgeons were provided with a notification identifying the issue and their responsibilities. These responsibilities include ensuring the affected personnel are aware of the contents of the notification. On 12/19/13, Distributors were sent a notification identifying the issue and to provide a copy of the notification sent to surgeons and risk managers. The distributors will also be instructed as needed to assist with identifying a comprehensive list of surgeons. " Communications outside of the United States will occur approximately two weeks after the United States communications. Effectiveness Checks Zimmer will conduct effectiveness checks in the following way: "Surgeons/Hospitals o100% of the FedEx notifications will be accounted for. oSecond and third notification attempts will be made. oFailed delivery attempts following the third attempt will result in the account being deemed as nonresponsive.
Quantity in commerce
5305 distribution events
Distribution
Worldwide Distribution. USA nationwide and Germany, with further distribution into EU countries.
Product codes and lots
Part numbers 00-5850-013-95

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0784-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0783-2014Class IITerminatedVoluntary: Firm initiated
Z-0784-2014Class IITerminatedVoluntary: Firm initiated