PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee...
DePuy Orthopaedics, Inc. began a recall of PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility on . FDA classified it as Class II and gives the reason as "Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1077-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- The recall is extended to the DePuy Distributor, Hospital, and Surgeon levels. The two affected distributors were contacted by telephone. Written communication to the two hospitals and one surgeon who received the devices will be delivered by DePuy Orthopaedics, Inc. via email or regular mail. The one affected surgeon will also be contacted by telephone by the Medical Safety Officer and a metallurgist, who will discuss the surgeon communication with the surgeon. The sales representatives aid customers in the affected device returns, as needed. The devices will be returned through the normal DePuy Returns process, to attention of Returns and marking H13-30C on the outside of the box. All (21) remaining devices have been verified as being returned and in quarantine. Effectiveness will be determined by product reconciliation, receipt of all hospital reconciliation forms, and receipt of international declaration from the International Affiliate. DePuy will follow-up with the affected hospitals until all hospital reconciliation forms are returned, and DePuy will follow-up with the International Affiliate until all international actions are complete and the international declaration is completed and return.
- Quantity in commerce
- 2
- Distribution
- Worldwide Distribution-USA including the states of IA, CA, KY, and FL, and the countries of Finland, Germany, Sweden, and Czech Republic.
- Product codes and lots
- Catalog Number 960015 Lot Numbers 7810268 and 7806929 (US) Lot Number 7806934 (OUS)