Z-0812-2014Class II

Vital Signs Anesthesia Circuits The Ventilator Circuit is intended for use as an interface between a...

Vital Signs Colorado, Inc. / initiated 2013-12-26 / Terminated / root cause: manufacturing process

Vital Signs Colorado Inc. began a recall of Vital Signs Anesthesia Circuits The Ventilator Circuit is intended for use as an interface between a ventilator dependent patient and a ventilator. The... on . FDA classified it as Class II and gives the reason as "A circuit in use was occluded by a blue port cap that was inadvertently attached to the circuit as a result of an internal inspection error and will cause occlusion of gas flow". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0812-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Vital Signs Colorado, Inc.
Product
Vital Signs Anesthesia Circuits The Ventilator Circuit is intended for use as an interface between a ventilator dependent patient and a ventilator. The Anesthesia Circuit is intended to administer medical gases to a patient during anesthesia. The CPAP Circuit is intended to increase the airway pressure of a patient.
Product code
CAH Filter, Bacterial, Breathing-Circuit
Reason for recall
A circuit in use was occluded by a blue port cap that was inadvertently attached to the circuit as a result of an internal inspection error and will cause occlusion of gas flow.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
GE Healthcare sent an Urgent Medical Device Correction letter on December 12, 2013 to all affected customers via certified letter to follow the safety instructions to: 1) Please follow all safety instructions in accordance with the instructions for use (IFU). 2) During installation, visually inspect the Vital Signs Breathing Circuits for any plugged ports. If 15mm or 22mm blue port caps are attached to any ports, please remove and discard. 3) Prior to use, the circuits should be pressure tested for leaks and flow tested for obstructions in accordance with the ventilator manufacturers specifications. The long-term solution is to update the instructions for the assembly operator and modify the circuit air test process by replacing the cap with a plug tethered to the workstation so the plug cannot be left on the circuit. If you have any questions or concerns regarding this notification, please contact GEMSIT Customer Service at +1-800-588-7044 (Domestic) or Vital Signs Customer Service at +1-800-932-0760 (International). Hours of Operation: 8:00 am EST to 6:00 pm EST. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.
Quantity in commerce
1,827,920 units US , 504,760 units OUS
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico; and Internationally to: Belgium, Canada, Chile, Mexico, Costa Rica, Kuwait, United Kingdom, Greece, Czech Republic, Latvia, Turkey, China, France, Italy, Brazil, Korea, Germany, Thailand, Egypt, Austria, Dominican Republic, India, Israel, Colombia, Japan, Netherlands, Venezula, Saudi Arabia, Argentina, Malaysia, Sweden, Lebanon, South Korea, Finland, United Arab Emirates, Taiwan, Panama, Uruguay, Singapore, Hong Kong, Ecuador, Peru, Oman, Spain, Lithuania, and South Africa.
Product codes and lots
PartNumber : 4S78XX00,155512,155518,156505,156545,1576162,1576214,542011,542021,542031,1553196,1572347,1553402,156205,156310,1574701,1574710,1574711,1574712,1574723,1572414,A4JXXXXX,6463,C311123X,C6X1XX1X,207002,C316123X,1560023,201814,1730327,20CF800C,C116123X,1570103,20CT8050,1577112,1552016,A4JXOXXX,CXX61X2X,1552017,1552018,20CT8048,20CT8055,M1171508,M1181639,1552020,M1162025,155200EU,1552017CN,1574792,M1171508VS,M1181639VS,1552026,1552027,1552028,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0812-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0812-2014Class IITerminatedVoluntary: Firm initiated