Vital Signs Anesthesia Circuits The Ventilator Circuit is intended for use as an interface between a...
Vital Signs Colorado Inc. began a recall of Vital Signs Anesthesia Circuits The Ventilator Circuit is intended for use as an interface between a ventilator dependent patient and a ventilator. The... on . FDA classified it as Class II and gives the reason as "A circuit in use was occluded by a blue port cap that was inadvertently attached to the circuit as a result of an internal inspection error and will cause occlusion of gas flow". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0812-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Vital Signs Colorado, Inc.
- Product
- Vital Signs Anesthesia Circuits The Ventilator Circuit is intended for use as an interface between a ventilator dependent patient and a ventilator. The Anesthesia Circuit is intended to administer medical gases to a patient during anesthesia. The CPAP Circuit is intended to increase the airway pressure of a patient.
- Product code
- CAH Filter, Bacterial, Breathing-Circuit
- Reason for recall
- A circuit in use was occluded by a blue port cap that was inadvertently attached to the circuit as a result of an internal inspection error and will cause occlusion of gas flow.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter on December 12, 2013 to all affected customers via certified letter to follow the safety instructions to: 1) Please follow all safety instructions in accordance with the instructions for use (IFU). 2) During installation, visually inspect the Vital Signs Breathing Circuits for any plugged ports. If 15mm or 22mm blue port caps are attached to any ports, please remove and discard. 3) Prior to use, the circuits should be pressure tested for leaks and flow tested for obstructions in accordance with the ventilator manufacturers specifications. The long-term solution is to update the instructions for the assembly operator and modify the circuit air test process by replacing the cap with a plug tethered to the workstation so the plug cannot be left on the circuit. If you have any questions or concerns regarding this notification, please contact GEMSIT Customer Service at +1-800-588-7044 (Domestic) or Vital Signs Customer Service at +1-800-932-0760 (International). Hours of Operation: 8:00 am EST to 6:00 pm EST. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.
- Quantity in commerce
- 1,827,920 units US , 504,760 units OUS
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico; and Internationally to: Belgium, Canada, Chile, Mexico, Costa Rica, Kuwait, United Kingdom, Greece, Czech Republic, Latvia, Turkey, China, France, Italy, Brazil, Korea, Germany, Thailand, Egypt, Austria, Dominican Republic, India, Israel, Colombia, Japan, Netherlands, Venezula, Saudi Arabia, Argentina, Malaysia, Sweden, Lebanon, South Korea, Finland, United Arab Emirates, Taiwan, Panama, Uruguay, Singapore, Hong Kong, Ecuador, Peru, Oman, Spain, Lithuania, and South Africa.
- Product codes and lots
- PartNumber : 4S78XX00,155512,155518,156505,156545,1576162,1576214,542011,542021,542031,1553196,1572347,1553402,156205,156310,1574701,1574710,1574711,1574712,1574723,1572414,A4JXXXXX,6463,C311123X,C6X1XX1X,207002,C316123X,1560023,201814,1730327,20CF800C,C116123X,1570103,20CT8050,1577112,1552016,A4JXOXXX,CXX61X2X,1552017,1552018,20CT8048,20CT8055,M1171508,M1181639,1552020,M1162025,155200EU,1552017CN,1574792,M1171508VS,M1181639VS,1552026,1552027,1552028,
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0812-2014 | Class II | Terminated | Voluntary: Firm initiated |