Z-1027-2014Class II
Pressfit Fem St Fx 16 X 1
Zimmer GmbH / initiated 2013-10-11 / Terminated / root cause: packaging
Zimmer, Inc. began a recall of PRESSFIT FEM ST FX 16 X 1 on . FDA classified it as Class II and gives the reason as "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1027-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Pressfit Fem St Fx 16 X 1
- Product code
- LWJ Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented
- Reason for recall
- Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
- Quantity in commerce
- 7,044,680 total
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
- Product codes and lots
- 60621428 60667768
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1001-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1002-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1003-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1004-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1005-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1006-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1007-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1008-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1009-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1010-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1011-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1012-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1013-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1014-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1015-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1016-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1017-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1018-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1019-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1020-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1021-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1022-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1023-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1024-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1025-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1026-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1027-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1028-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1029-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1030-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1031-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1032-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1033-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1034-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1035-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1036-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1037-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1038-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1039-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1040-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1041-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1042-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1043-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1044-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.