Z-0779-2014Class II

Anspach MIA 16 (16cm Minimally Invasive Attachment), Rx Only MIA16 attachments are intended for use with...

The Anspach Effort, Inc. / initiated 2013-05-06 / Terminated / root cause: sterility

The Anspach Effort, Inc. began a recall of Anspach MIA 16 (16cm Minimally Invasive Attachment), Rx Only MIA16 attachments are intended for use with Motor Systems for cutting and shaping bones, including... on . FDA classified it as Class II and gives the reason as "Product did not reach expected sterility assurance level". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0779-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Anspach Effort, Inc.
Product
Anspach MIA 16 (16cm Minimally Invasive Attachment), Rx Only MIA16 attachments are intended for use with Motor Systems for cutting and shaping bones, including the spine and cranium.
Product code
ERL Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece
Reason for recall
Product did not reach expected sterility assurance level
Root cause tag
sterility
FDA root cause
Unknown/Undetermined by firm
Action
The firm, Aspach, sent an "URGENT: MEDICAL DEVICE LABELING CORRECTION" letter dated May 13, 2013 to its consignees/customers. The letter described the product, problem and actions to be taken. The consignees/ customers were instructed that Sterrad 100S system should not be used on the MIA 16 attachment (the information only applies to only Sterrad sterilization of this devices and not to any other attachments.), and complete and return the Return Receipt Requested Customer Reply form via fax to: 1-800-327-6661 or email: [email protected]. If you distribute any of the products to other services or facilities, please forward this information as appropriate. Should you have any queries, please do not hesitated to contact Anspach Product Support at (800) 327-6887. Hours of Operation are 8:30am-5:30pm Eastern Standard Time or email: [email protected].
Quantity in commerce
183
Distribution
Nationwide Distribution: FL, IL, CA. IN, MI, CA, NJ, TX, NE, NC, OR, PA, TN, DC, UT, CO, MO, KY, MD, WV, NY, and MN.
Product codes and lots
ALL LOTS

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0779-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0779-2014Class IITerminatedVoluntary: Firm initiated