Z-1167-2014Class II

Navigator Surgical Kit

ZimVie US Corp LLC / initiated 2013-07-22 / Terminated / root cause: sterility

Biomet 3i, LLC began a recall of Navigator Surgical Kit Product Usage: Molded plastic tray used in holding various dental instruments for autoclave processing on . FDA classified it as Class II and gives the reason as "During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1167-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Navigator Surgical Kit Product Usage: Molded plastic tray used in holding various dental instruments for autoclave processing.
Product code
LRO General Surgery Tray
Reason for recall
During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required.
Root cause tag
sterility
FDA root cause
Unknown/Undetermined by firm
Action
Biomet 3i sent an Urgent Medical Device Field Correction Notice dated July 2013 to all affected customer via email, facsimile, or postal mail. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to complete the attached response form and fax to 561-514-6316. For questions contact Biomet 3i customer service at 1-800-342-5454.
Quantity in commerce
21
Distribution
Worldwide Distribution - USA Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, including Hawaii and Puerto Rico and the countries of : Argentina, Austria, Canada, China, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain, Hong Kong, Hungary, Israel, Japan, Korea, Mexico, Poland, Portugal, Paraguay, Sweden, Singapore, Taiwan, and Uruguay.
Product codes and lots
SGKIT (SGTRAY)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1167-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1167-2014Class IITerminatedVoluntary: Firm initiated
Z-1168-2014Class IITerminatedVoluntary: Firm initiated
Z-1169-2014Class IITerminatedVoluntary: Firm initiated
Z-1170-2014Class IITerminatedVoluntary: Firm initiated
Z-1171-2014Class IITerminatedVoluntary: Firm initiated