Z-1083-2014Class II

Integrity. Version 1.1 of the Integrity interface and control software for the Elekta range of medical...

Elekta Limited / initiated 2013-11-25 / Terminated / root cause: software

Elekta, Inc. began a recall of Integrity. Version 1.1 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators on . FDA classified it as Class II and gives the reason as "During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1083-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Integrity. Version 1.1 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Root cause tag
software
FDA root cause
Software design
Action
Corrective Action #1 - Release of Important Field Safety Notice 200 01 502 043 - "Recommended quality assurance after a calibration procedure" Released 25 November, 2013. Corrective Action #2 - Permanent Solution - All users of Integrity 3.1 software will be informed or a mandatory upgrade to Integrity 3.2 upon its release which will prevent calibration from being inverted. Target Release for Integrity 3.2 will be available at the end of February 2014.
Quantity in commerce
455
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CT, FL, GA, IL, KY, LA, MD, MA, MI, MO, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, and WY, and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Czech Republic, Cuba, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan , Panama, Peru, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Ukraine, Venezuela, and Vietnam.
Product codes and lots
Model 1.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1083-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1083-2014Class IITerminatedVoluntary: Firm initiated
Z-1084-2014Class IITerminatedVoluntary: Firm initiated
Z-1085-2014Class IITerminatedVoluntary: Firm initiated