VirtuoSaph Plus Endoscopic Vessel Harvesting System, Part number: VSP550, Sterile***Rx Only***
Terumo Cardiovascular Systems Corporation began a recall of VirtuoSaph Plus Endoscopic Vessel Harvesting System, Part number: VSP550, Sterile***Rx Only*** Product Usage: The VirtuoSaph Plus Endoscopic Vessel Harvesting... on . FDA classified it as Class II and gives the reason as "Sterility of medical devices intended for use in surgical procedures may be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1069-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- VirtuoSaph Plus Endoscopic Vessel Harvesting System, Part number: VSP550, Sterile***Rx Only*** Product Usage: The VirtuoSaph Plus Endoscopic Vessel Harvesting System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Sterility of medical devices intended for use in surgical procedures may be compromised.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Terumo CVS notified afected end users by phone on December 12, 2013 and advised that they had become aware that the sterile packaging barrier was breached in specific lots of VirtuoSaph Plus Endoscopic Vessel Harvesting Systems. Customers were requested to return all unused affected product and verify that all users at their institution have been made aware of the risks associated with using the affected devices. For questions call 734-663-4145.
- Quantity in commerce
- 26 units
- Distribution
- USA Nationwide Distribution in the state of LA, NJ, and NY.
- Product codes and lots
- Lot numbers: 21K, 31K, 32K, 33K, 34K, 35K
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1069-2014 | Class II | Terminated | Voluntary: Firm initiated |