Z-1081-2014Class II

RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as...

Elekta Limited / initiated 2013-11-25 / Terminated / root cause: software

Elekta, Inc. began a recall of RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear... on . FDA classified it as Class II and gives the reason as "During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1081-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Root cause tag
software
FDA root cause
Software design
Action
Elekta sent an Important Field Safety Notice November 2013 to all affected customers. Corrective Action #1 - Release of Important Field Safety Notice 200 01 502 041 - "Recommended quality assurance after a calibration procedure" Released 25 November, 2013. Important Field Safety Notice is in the process of being distributed to customers advising them of the recommended Quality Assurance checks to avoid/detect the problem. The permanent solution to this problem is currently been developed. Corrective Action #2 - Permanent Solution - All users of Desktop Pro software will be informed of a mandatory upgrade to Integrity 1.2 upon its release which will prevent calibration from being inverted. For questions regarding this recall call 770-300-9725.
Quantity in commerce
489
Distribution
Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.
Product codes and lots
R4.2, R5.0, R5.1, R6.0, R6.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1081-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1081-2014Class IITerminatedVoluntary: Firm initiated