XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation...
Elekta, Inc. began a recall of XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed on . FDA classified it as Class II and gives the reason as "When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0928-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Corrective Action 1 - An important safety notice, LINXIO0018 - Incorrect Effective Depth for Bolused Beams is in the process of being distributed to affected customers. (Release date November 25, 2013). Corrective Action 2 - Develop a fix to the defect in an expeditious timeframe and make available in Patch Release 4.0.03 and Release 5.00.01.
- Quantity in commerce
- 1424
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, QA, WA, WV, WI, and WY, and the countries of Albania, Algeria, Argentina, Austria, Australia, Bosnia, Bulgaria, Belgium, Bahrain, Brazil, Belarus, Botswana, Bangladesh, Canada, China, Columbia, Costa Rica, Croatia, Cuba, Cyprus, Chile, Czech Republic, Ecuador, Egypt, Estonia, France, Finland, Gabon, Germany, Great Britain, Greece, Guatemala, Hungary, India, Indonesia, Italy, Ireland, Israel, Iraq, Jordan, Japan, Kosovo, Kuwait, Lithuania, Libya, Latvia, Morocco, Malaysia, Malta, Mexico, Myanmar, Netherland, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sudan, Syria, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, Venezuela, and Vietnam.
- Product codes and lots
- MUJ
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0928-2014 | Class II | Terminated | Voluntary: Firm initiated |