Stryker Orthopaedics Navigation Compatible Accolade Broach Handle 1 single unit to a package Howmedica...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Navigation Compatible Accolade Broach Handle 1 single unit to a package Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA... on . FDA classified it as Class II and gives the reason as "Stryker received reports from the field of the impaction plate dissociating/fracturing from the main body of the broach handle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0964-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Navigation Compatible Accolade Broach Handle 1 single unit to a package Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France Made in USA non-sterile
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Stryker received reports from the field of the impaction plate dissociating/fracturing from the main body of the broach handle.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Stryker Orthopaedics sent notification letters and acknowledgement forms via Fed Ex as follows: Branches-11/20/2013 Hospital Risk Manager - 11/21/2013
- Quantity in commerce
- 211 units
- Distribution
- IL MA, MN, ND, OR, & PA
- Product codes and lots
- Catalog no. 2124-1400 All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0964-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.