Z-0937-2014Class II

Hemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used in...

Accriva Diagnostics, Inc. / initiated 2013-12-19 / Terminated / root cause: design

ITC-Nexus Dx began a recall of Hemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used in performing a quantitative, one -stage APTT. The... on . FDA classified it as Class II and gives the reason as "ITC has determined that some Citrate APTT cuvettes may recover higher than expected results in normal individuals". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0937-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accriva Diagnostics, Inc.
Product
Hemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used in performing a quantitative, one -stage APTT. The citrate APTT test is used for evaluation of low doses of heparin anticoagulant (up to 1.5 units/mL), depending upon individual patient heparin sensitivity.
Product code
GFO Activated Partial Thromboplastin
Reason for recall
ITC has determined that some Citrate APTT cuvettes may recover higher than expected results in normal individuals.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, ITC The Point of Care, sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 19, 2013 a long with an attached Health Professional Account Tracking Form via UPS to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to forward this communication to all those within your organization who need to be aware of this matter including any facility that you forwarded any of the affected product to; check all inventory to determine if you have any of the affected lots listed; if you have any affected lots, stop using them, remove from inventory and return to ITC; complete the attached Health Professional Account Tracking Form and return it within 10 days via fax to: 866-429-3132, email: [email protected] or mail to: ITC Technical Support, 20 Corporate Place South - Piscataway, NJ 08854 . It is important that you complete the form whether or not you have remaining inventory of these lots. In addition, contact ITC Technical Support immediately if you become aware of an unreported adverse event related to the use of affected J103C cuvettes. ITC will review your response form and will contact you to arrange for any inventory to be returned to ITC and ITC will replace all unused, returned product. If you have questions, please contact ITC Technical Support: Telephone 800-579-2255 or 732-548-5700, Extension 4011 or E-mail [email protected].
Quantity in commerce
25,965 curvettes (7,425 cuvettes US; 18,540 cuvettes OUS)
Distribution
Worldwide Distribution: US (nationwide) and countries of: Austria, Belgium, Canada, Serbia, Germany, Estonia, Italy, Spain and Saudia Arabia.
Product codes and lots
E3JCC013-P1 - exp 2014-05 E3JCC014-P2 - exp - 2014-05 F3JCC015-P3 - exp -2014-06 F3JCC016 - exp - 2014-06 H3JCC017 - exp - 2014-08

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0937-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0937-2014Class IITerminatedVoluntary: Firm initiated