Z-0870-2014Class II

BD BACTEC FX 40 instrument, catalog number 442296, manufactured by BD Diagnostic Systems, 7 Loveton Circle...

Becton Dickinson & Co / initiated 2013-12-09 / Terminated / root cause: manufacturing process

Becton Dickinson & Co. began a recall of BD BACTEC FX 40 instrument, catalog number 442296, manufactured by BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD 21152. The instrument is designed for... on . FDA classified it as Class II and gives the reason as "Improperly functioning component of a diagnostic medical device may cause false negative results in clinical specimens". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0870-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD BACTEC FX 40 instrument, catalog number 442296, manufactured by BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD 21152. The instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.
Product code
MDB System, Blood Culturing
Reason for recall
Improperly functioning component of a diagnostic medical device may cause false negative results in clinical specimens.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
BD Diagnostics initiated a field correction by phone on December 9, 2013, to affected customers and followed the calls with letters, e-mails, or fax. BD Diagnostics sent a secondary Urgent Field Corrective Action letter dated November 2013 to all affected customers. The recalling firm Secondary notifications were serial number specific and explained that some BD BACTEC FX instruments contain racks that may demonstrate variable indicator light intensities. Specifically, the Anonymous vial indicator lights which should appear yellow may instead appear more green in color. BD noted that affected instruments continue to function properly and will not report an incorrect result. The issue is limited to the yellow indicator lights only. Red and green indicator lights appear as intended. BD is developing software to correct this issue and anticipates that the new software will be available to begin instrument updates by February, 2014. The new software will not affect instrument if the indicator lights are functioning as intended. Until the software can be updated on your instrument, BD provided workflow instructions to ensure that the indicator light issue does not occur. For questions regarding this recall call 410-316-4054.
Quantity in commerce
2 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to IN, AU, CN, AR, BE, PE, KR, BR, CO, SG, TH, BN, HK, JP, and CA.
Product codes and lots
Catalog number 442296

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0870-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0869-2014Class IITerminatedVoluntary: Firm initiated
Z-0870-2014Class IITerminatedVoluntary: Firm initiated