510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1057-2014Class II

Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their...

Synthes GmbH / initiated 2014-01-07 / Terminated / root cause: manufacturing process

Synthes, Inc. began a recall of Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving... on . FDA classified it as Class II and gives the reason as "The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1057-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Synthes sent an Urgent Notice - Medical Device Recall Letter dated January 7 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Action to be taken: Our records indicate that your facility has the product(s) subject to this Recall. Synthes asks that you review your inventory and immediately remove the affected lots from stock. Please take the following actions: If you DO have any of the identified devices, please take the following steps: "Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization Number. "Complete the Verification Section at the end of this letter by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, telephone number and signature in the spaces provided. "Return the Verification Form (page-3 of this letter) with the product to: "Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132. If you DO NOT have the identified product, please take the following steps: "Complete the attached Verification Section at the end of this letter by checking the appropriate box indication that no affected product has been located. :Please include your name, title, telephone number and signature in the spaces provided this return documentation acknowledges your receipt of medical device removal information. "Return the documents to DePuy Synthes by: " Fax: 877-271-1473 or " Scan/email: [email protected] If you have any questions,please call 610-719-5450. Thank you for your attention to this issue.
Quantity in commerce
143
Distribution
USA (nationwide) Distribution.
Product codes and lots
Part No. 03.010.523, Lot Nos. 8279162 & 8429004

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1057-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1057-2014Class IITerminatedVoluntary: Firm initiated