Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures...
Synthes, Inc. began a recall of Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small... on . FDA classified it as Class II and gives the reason as "The Drill Template and the Saw Guide for the Ulna Osteotomy System may exhibit jamming, bending, or breaking of the screw connecting the Saw Guide to the Drill Template or jamming, bending, or breaking of the K-wire...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1274-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, including osteopenic bone in the ulna.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The Drill Template and the Saw Guide for the Ulna Osteotomy System may exhibit jamming, bending, or breaking of the screw connecting the Saw Guide to the Drill Template or jamming, bending, or breaking of the K-wire during fixation of the Drill Template.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- A recall notification letter, dated January 7, 2013, was sent to sales consultants and Users.
- Quantity in commerce
- 343
- Distribution
- Nationwide Distribution-including the states of AR, AK, VA, IN, PA, OH, CT, TX, MI, WI, NC, CO, IL, and MA.
- Product codes and lots
- Part Nos. 03.111.900 through 03.111.905, with Lot Nos. 779372, 2638601, 7769386, 2638603, 7836420, 7772531, 2638604, 7836422, 7772538, 2638605, 7836423, 7915554, 7772543, 7772542, 7872775, 8400287, 7772554, 7772555, and 8229040.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1274-2014 | Class II | Terminated | Voluntary: Firm initiated |