T:slim Insulin Delivery System Insulin Infusion Pump
Tandem Diabetes Care Inc began a recall of T:slim Insulin Delivery System Insulin Infusion Pump on . FDA classified it as Class I and gives the reason as "Specific lots of insulin cartridges that are used with the T:slim Insulin Pump may be at risk for leaking. A cartridge leak could result in the device delivering too much or too little insulin, which can lead to a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0827-2014
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tandem Diabetes Care, Inc.
- Product
- T:slim Insulin Delivery System Insulin Infusion Pump
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- Specific lots of insulin cartridges that are used with the T:slim Insulin Pump may be at risk for leaking. A cartridge leak could result in the device delivering too much or too little insulin, which can lead to a serious adverse event.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Tandem Diabetes Care sent an Urgent Medical Device Voluntary Recall letter dated January 10, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed check the lot numbers of their cartridge supply against the llist of affected lot numbers. Do Not use any cartridges from the lots affected by this recall. If customers don't have useable cartridges, discontinue using the pump and revert to their backup plan until new supplies arrive. Customers were instructed to call Tandem Technical Support at 1-877-801-6901 to receive replacement cartridges at no charge. Customers were asked to return the enclosed return response form via email, fax, or mail. UPDATED: Tandem Diabetes Care expanded the recall om January 20, 2014. Tandem Diabetes Care, Inc. (NASDAQ: TNDM), or Tandem, today announced that it is expanding a voluntary recall of specific lots of insulin cartridges that are used with the t:slim Insulin Pump. For questions regarding this recall calll 858-366-6900, ext 6963.
- Quantity in commerce
- 12, 807 boxes (10packs)
- Distribution
- USA (nationwide)
- Product codes and lots
- Lot Numbers: M000857, M001414, M001454, M001963, M002028, M000869, M001415, M001455, M001964, M002029, M001344, M001416, M001456, M001973, M002030, M001345, M001417, M001457, M001974, M002082, M001346, M001420, M001458, M001979, M002083, M001347, M001421, M001459, M001980, M002096, M001389, M001422, M001460, M001987, M002097, M001390, M001423, M001528, M001988, M002099, M001391, M001451, M001529, M001990, M002100, M001392, M001452, M001530, M001991, M002119, M001393, M001453, M001532, M002027, M002120.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0827-2014 | Class I | Terminated | Voluntary: Firm initiated |