Z-1062-2014Class II
Bausch & Lomb AKREOS AO Micro Incision Lens The Akreos IOL intended for primary implantation in the capsular...
Bausch & Lomb Incorporated / initiated 2013-10-16 / Terminated / root cause: manufacturing process
Bausch & Lomb Surgical, Inc. began a recall of Bausch & Lomb AKREOS AO Micro Incision Lens The Akreos IOL intended for primary implantation in the capsular bag of the eye for the visual correction of... on . FDA classified it as Class II and gives the reason as "Lens was manufactured with incorrect raw material". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1062-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bausch & Lomb Incorporated
- Product
- Bausch & Lomb AKREOS AO Micro Incision Lens The Akreos IOL intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia secondary to the removal of a cataractous lens in adult patients.
- Product code
- HQL Intraocular Lens
- Reason for recall
- Lens was manufactured with incorrect raw material.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Bausch + Lomb, telephoned and sent an "URGENT - MEDICAL DEVICE RECALL" letter dated October 17, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to: 1) Determine the disposition of the lenses; 2) Complete and provide the enclosed acknowledgement form to the sales respresentatives collecting the lenses, and 3) Return all unused products. If you have any questions, please contact Bausch + Lomb at (800) 338-2020.
- Quantity in commerce
- 336 IOLs (283 IOLs in the US, 53 IOLs outside the US)
- Distribution
- Worldwide Distribution: US (nationwide) and Internationally to: Great Britain, France, Spain, Portugal, Sweden, Russia and Guadeloupe.
- Product codes and lots
- Model Numbers(s): AO60 and MI60L