Z-0900-2014Class II

Atlantis PV, 8.5F, 15 MHz Peripheral Imaging Catheter: Sterile, R; Sterilized using irradiation

Boston Scientific Corporation / initiated 2014-01-06 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Atlantis PV, 8.5F, 15 MHz Peripheral Imaging Catheter: Sterile, R; Sterilized using irradiation. Product Usage: This catheter is a 15 MHz ultrasound imaging... on . FDA classified it as Class II and gives the reason as "Reports of physicians experiencing inability to pass a 0.035" guidewire through the guidewire lumen and exit port at the (proximal) Y-manifold of the imaging catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0900-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Atlantis PV, 8.5F, 15 MHz Peripheral Imaging Catheter: Sterile, R; Sterilized using irradiation. Product Usage: This catheter is a 15 MHz ultrasound imaging catheter intended to operate with an IVUS instrument for diagnostic imaging. It is used with an 8.5F-introducer sheath and a 0.035 guidewire.
Product code
ITX Transducer, Ultrasonic, Diagnostic
Reason for recall
Reports of physicians experiencing inability to pass a 0.035" guidewire through the guidewire lumen and exit port at the (proximal) Y-manifold of the imaging catheter.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Boston Scientific sent an Urgent Medical Device Recall Removal letter dated January 14, 2014 to all affected customers by overnight mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete an Account Reply verification Tracking Form in order to received a Return Goods Authorization number for product return. For questions contact your local Sales Representative.
Quantity in commerce
1068 single units
Distribution
US Nationwide Distribution and Mexico
Product codes and lots
Product number: H749364560, Catalog number: 36456; Lot numbers:  16159358, 16183964, 16230195, 16230673, 16247263, 16257809, 16268498, 16289177, 16311356, 16330543, 16342824, 16419885, 16420624, 16420903, 16447744, 16458536, 16486133, 16494530; Expiration Date: June 11th 2014 to October 29th 2014.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0900-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0900-2014Class IITerminatedVoluntary: Firm initiated