Z-0979-2014Class II

GEMINI TF Big Bore PET/CT, Computer Tomography System, Philips Healthcare System, Cleveland, OH

Philips North America LLC / initiated 2013-12-05 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of GEMINI TF Big Bore PET/CT, Computer Tomography System, Philips Healthcare System, Cleveland, OH on . FDA classified it as Class II and gives the reason as "The TF Big Bore patient table experienced mechanical binding during horizontal table motion resulting in an automatic Emergency Stop (E-stop). An E-Stop will interrupt a current scan, resulting in an incomplete study". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0979-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
GEMINI TF Big Bore PET/CT, Computer Tomography System, Philips Healthcare System, Cleveland, OH
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
The TF Big Bore patient table experienced mechanical binding during horizontal table motion resulting in an automatic Emergency Stop (E-stop). An E-Stop will interrupt a current scan, resulting in an incomplete study.
Root cause tag
software
FDA root cause
Software design
Action
ON 12/5/2013 the firm sent Medical Device Correction Letters to their customers.
Quantity in commerce
47 Units
Distribution
Product was shipped to the following states: AR, CA, CO, FL, IL, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, PA, TX, VA, VT & WI. Product was also shipped to the following countries: Belgium, Canada, China, Denmark, France, Germany, Italy, Japan, Netherlands, Saudi Arabia & Venezuela.
Product codes and lots
Model #882476, Serial #'s: 9004, 9005, 9006, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9206, 9207, 9208, 9209, 9210, 9211, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, 9221, 9222, 9223, 9224, 9225, 9226, 9228, 9229, 9231.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0979-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0979-2014Class IITerminatedVoluntary: Firm initiated