Z-0832-2014Class II

Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The...

Philips North America LLC / initiated 2013-12-09 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed... on . FDA classified it as Class II and gives the reason as "Patient images exhibited ring artifacts". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0832-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Patient images exhibited ring artifacts.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Philips Healthcare, sent an "URGENT - Medical Device Correction" letter dated December 9, 2013 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following to avoid the occurrence of these specific ring artifacts: 1. Disable Automatic Scan Time for all Exam Cards 2. Leave Automatic Scan Time on during planning and prior to clicking Go to initiate the scan, click the acquisition series of the scan and disable Automatic Scan Time, then proceed with the scan. 3. Disable Edit Before Final Recon (Reconstruction) for all Exam Cards If a ring artifact is present, clinical judgment should be used by the customer/user to determine if the images acquired in the dataset contain enough information to make a diagnosis. Philips will be releasing a Field Change Order (FCO) to correct the issue with software version release 4.1.1.19261. In addition, a Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Quantity in commerce
7 units
Distribution
Worldwide Distribution: US (nationwide) including states of: FL, IN, NY, OR and PA; and internationally to: Australia and Israel.
Product codes and lots
Brilliance iCT: 728306, Serial #'s: 100018, 100040, 100411, 100503, 100506  Brilliance iCT SP: 728311, Serial #'s: 200047, 200121

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0832-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0832-2014Class IITerminatedVoluntary: Firm initiated