510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1103-2014Class II

Optiflux F200NR High Flux Dialyzers; Product Number: 0500320N. Single Use acute or chronic hemanalysis...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2014-02-04 / Terminated

Fresenius Medical Care Holdings, Inc. began a recall of Optiflux F200NR High Flux Dialyzers; Product Number: 0500320N. Single Use acute or chronic hemanalysis treatment on . FDA classified it as Class II and gives the reason as "Dialysate Port Leak During Priming of the Dialyzers". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1103-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Optiflux F200NR High Flux Dialyzers; Product Number: 0500320N. Single Use acute or chronic hemanalysis treatment.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Dialysate Port Leak During Priming of the Dialyzers
FDA root cause
Under Investigation by firm
Action
Fresenius Medical issued Notice to Clinics and Clinics with Patients Letters Dated - 1st Mailing - February 4, 2014 and Patients Letters Dated - February 5, 2014 - these letters contained the correct information.Corrected letters to Clinics and Clinics with Patients issued 2/14/14 via certified mail with return receipt and fax-back form . Users are advised the leak may be stopped by tightening the dialysate port cap more securely. Questions associated with this product notification to your Customer Service Team at 1-800-323-5188.
Quantity in commerce
19,596 units
Distribution
Worldwide distribution: USA (nationwide) and the country of: Mexico.
Product codes and lots
Lot Number: 13NU04008

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1103-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1098-2014Class IITerminatedVoluntary: Firm initiated
Z-1099-2014Class IITerminatedVoluntary: Firm initiated
Z-1100-2014Class IITerminatedVoluntary: Firm initiated
Z-1101-2014Class IITerminatedVoluntary: Firm initiated
Z-1102-2014Class IITerminatedVoluntary: Firm initiated
Z-1103-2014Class IITerminatedVoluntary: Firm initiated
Z-1104-2014Class IITerminatedVoluntary: Firm initiated
Z-1105-2014Class IITerminatedVoluntary: Firm initiated
Z-1106-2014Class IITerminatedVoluntary: Firm initiated