Z-1105-2014Class II
Optiflux B200 Flux Dialyzers; Product Number: 0500325E. Single Use acute or chronic hemanalysis treatment
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2014-02-04 / Terminated
Fresenius Medical Care Holdings, Inc. began a recall of Optiflux B200 Flux Dialyzers; Product Number: 0500325E. Single Use acute or chronic hemanalysis treatment on . FDA classified it as Class II and gives the reason as "Dialysate Port Leak During Priming of the Dialyzers". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1105-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Optiflux B200 Flux Dialyzers; Product Number: 0500325E. Single Use acute or chronic hemanalysis treatment.
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Dialysate Port Leak During Priming of the Dialyzers
- FDA root cause
- Under Investigation by firm
- Action
- Fresenius Medical issued Notice to Clinics and Clinics with Patients Letters Dated - 1st Mailing - February 4, 2014 and Patients Letters Dated - February 5, 2014 - these letters contained the correct information.Corrected letters to Clinics and Clinics with Patients issued 2/14/14 via certified mail with return receipt and fax-back form . Users are advised the leak may be stopped by tightening the dialysate port cap more securely. Questions associated with this product notification to your Customer Service Team at 1-800-323-5188.
- Quantity in commerce
- 100,008 units
- Distribution
- Worldwide distribution: USA (nationwide) and the country of: Mexico.
- Product codes and lots
- Lot Numbers: 13LU02011 13LU02013 13NU02016 13NU02017
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1098-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1099-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1100-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1101-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1102-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1103-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1104-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1105-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1106-2014 | Class II | Terminated | Voluntary: Firm initiated |