Z-0845-2014Class II

InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V)

Philips North America LLC / initiated 2014-01-13 / Terminated / root cause: manufacturing process

Philips Healthcare began a recall of InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended... on . FDA classified it as Class II and gives the reason as "Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0845-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908
Product code
NCX System, Hypothermia, Intravenous, Cooling
Reason for recall
Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Philips Healthcare Recall Notice for this proactive performance correction was provided verbally to hospital biomedical engineers. Philips product support engineering scheduled and conducted the update on customer devices to the most current MCB configuration.
Quantity in commerce
50
Distribution
Worldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.
Product codes and lots
Model Numbers: 861470 (120V), 861472 (240V).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0845-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0845-2014Class IITerminatedVoluntary: Firm initiated