InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V)
Philips Healthcare began a recall of InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended... on . FDA classified it as Class II and gives the reason as "Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0845-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908
- Product code
- NCX System, Hypothermia, Intravenous, Cooling
- Reason for recall
- Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Philips Healthcare Recall Notice for this proactive performance correction was provided verbally to hospital biomedical engineers. Philips product support engineering scheduled and conducted the update on customer devices to the most current MCB configuration.
- Quantity in commerce
- 50
- Distribution
- Worldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.
- Product codes and lots
- Model Numbers: 861470 (120V), 861472 (240V).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0845-2014 | Class II | Terminated | Voluntary: Firm initiated |