Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart...
Edwards Lifesciences, LLC began a recall of Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation on . FDA classified it as Class II and gives the reason as "Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0918-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.
- Product code
- NPT Aortic Valve, Prosthesis, Percutaneously Delivered
- Reason for recall
- Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Edwards Lifesciences issued an Urgent Product Notice to all customers. Letters were sent via Fed Ex on 01/20/14. The letter outlines steps to be taken to enable a valve to fit into the open orifice, and allow the valve to be crimped according to normal instructions. Customers were instructed to forward the Urgent Product Notification letter to necessary personnel within their organization. Customers were also instructed to acknowledge they have reviewed and understand the notification letter by signing and dating the form included with the letter. Customers with questions were instructed to contact their Edwards Clinical Specialist.
- Quantity in commerce
- 422
- Distribution
- Nationwide Distribution
- Product codes and lots
- lot 59617777 and 59658495.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0918-2014 | Class II | Terminated | Voluntary: Firm initiated |