Z-0918-2014Class II

Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart...

Edwards Lifesciences LLC / initiated 2014-01-20 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation on . FDA classified it as Class II and gives the reason as "Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0918-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.
Product code
NPT Aortic Valve, Prosthesis, Percutaneously Delivered
Reason for recall
Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Edwards Lifesciences issued an Urgent Product Notice to all customers. Letters were sent via Fed Ex on 01/20/14. The letter outlines steps to be taken to enable a valve to fit into the open orifice, and allow the valve to be crimped according to normal instructions. Customers were instructed to forward the Urgent Product Notification letter to necessary personnel within their organization. Customers were also instructed to acknowledge they have reviewed and understand the notification letter by signing and dating the form included with the letter. Customers with questions were instructed to contact their Edwards Clinical Specialist.
Quantity in commerce
422
Distribution
Nationwide Distribution
Product codes and lots
lot 59617777 and 59658495.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0918-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0918-2014Class IITerminatedVoluntary: Firm initiated