Z-0973-2014Class II

Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging...

Medtronic Navigation, Inc / initiated 2013-07-25 / Terminated / root cause: labeling and UDI

Medtronic Navigation, Inc. began a recall of Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D... on . FDA classified it as Class II and gives the reason as "Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0973-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.
Product code
OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Reason for recall
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
Root cause tag
labeling and UDI
FDA root cause
Radiation Control for Health and Safety Act
Action
Medtronic Navigation Inc. issued Notiifcation on 3/4/14 and included thee errata sheet to all affected customers updating the missing and incorrect information and ensure that labeling is now in compliance with 21 CFR Part 1020.30(h). This will be followed by a software update that contains the same information on the errata sheet, which serves to correct the electronic version of the Instructions for Use. The paper errata sheet will be sent via traceable means and the software upgrade will be performed by Medtronic Service either through remote access or by site visit of Medtronic service personnel. For further questions please call (978) 698-6000.
Quantity in commerce
449 in US
Distribution
USA (nationwide) Distribution
Product codes and lots
Model No. BI-700-00027 and BI-700-00028 Serial number range: 00102 through 00753 and B13080721, B13080722, B13090729, B13120763, B13120764, B13120766, B13120768, B13120769

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0973-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0973-2014Class IITerminatedVoluntary: Firm initiated