Z-1041-2014Class II

2.7 X 6 Cort Screw/selfta 2.7 X 7 Cort Screw/selfta 2.7 X 8 Cort Screw/selfta 2.7 X 9 Cort Screw/selfta 2.7 X...

Zimmer GmbH / initiated 2013-10-11 / Terminated / root cause: packaging

Zimmer, Inc. began a recall of 2.7 X 6 CORT SCREW/SELFTA 2.7 X 7 CORT SCREW/SELFTA 2.7 X 8 CORT SCREW/SELFTA 2.7 X 9 CORT SCREW/SELFTA 2.7 X 10 CORT SCREW/SELFT 2.7 X 12 CORT SCREW/SELFT 2.7... on . FDA classified it as Class II and gives the reason as "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1041-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
2.7 X 6 Cort Screw/selfta 2.7 X 7 Cort Screw/selfta 2.7 X 8 Cort Screw/selfta 2.7 X 9 Cort Screw/selfta 2.7 X 10 Cort Screw/selft 2.7 X 12 Cort Screw/selft 2.7 X 14 Cort Screw/selft 2.7 X 16 Cort Screw/selft 2.7 X 18 Cort Screw/selft 2.7 X 20 Cort Screw/selft 2.7 X 22 Cort Screw/selft 2.7 X 24 Cort Screw/selft 2.7 X 26 Cort Screw/selft 2.7 X 28 Cort Screw/selft 2.7 X 30 Cort Screw/selft 2.7 X 32 Cort Screw/selft 2.7 X 34 Cort Screw/selft 2.7 X 36 Cort Screw/selft 2.7 X 38 Cort Screw/selft 2.7 X 40 Cort Screw/selft 2.7 X 42 Cort Screw/selft 2.7 X 44 Cort Screw/selft 2.7 X 45 Cort Screw/selft 2.7 X 46 Cort Screw/selft 2.7 X 48 Cort Screw/selft 2.7 X 50 Cort Screw/selft 4.0 X 20 Canc Screw/selft
Product code
HWC Screw, Fixation, Bone
Reason for recall
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Root cause tag
packaging
FDA root cause
Packaging
Action
Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Quantity in commerce
7,044,680 total
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Product codes and lots
61173335 61173337 61173339 61173340 61173341 61115956 61173344 61173346 61173347 61173349 61173350 61173352 61173353 61173354 61173355 61173356 61173357 61173358 61173359 61173360 61173361 61173362 61173363 61173364 61173365 61173366 61359816

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1041-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1001-2014Class IITerminatedVoluntary: Firm initiated
Z-1002-2014Class IITerminatedVoluntary: Firm initiated
Z-1003-2014Class IITerminatedVoluntary: Firm initiated
Z-1004-2014Class IITerminatedVoluntary: Firm initiated
Z-1005-2014Class IITerminatedVoluntary: Firm initiated
Z-1006-2014Class IITerminatedVoluntary: Firm initiated
Z-1007-2014Class IITerminatedVoluntary: Firm initiated
Z-1008-2014Class IITerminatedVoluntary: Firm initiated
Z-1009-2014Class IITerminatedVoluntary: Firm initiated
Z-1010-2014Class IITerminatedVoluntary: Firm initiated
Z-1011-2014Class IITerminatedVoluntary: Firm initiated
Z-1012-2014Class IITerminatedVoluntary: Firm initiated
Z-1013-2014Class IITerminatedVoluntary: Firm initiated
Z-1014-2014Class IITerminatedVoluntary: Firm initiated
Z-1015-2014Class IITerminatedVoluntary: Firm initiated
Z-1016-2014Class IITerminatedVoluntary: Firm initiated
Z-1017-2014Class IITerminatedVoluntary: Firm initiated
Z-1018-2014Class IITerminatedVoluntary: Firm initiated
Z-1019-2014Class IITerminatedVoluntary: Firm initiated
Z-1020-2014Class IITerminatedVoluntary: Firm initiated
Z-1021-2014Class IITerminatedVoluntary: Firm initiated
Z-1022-2014Class IITerminatedVoluntary: Firm initiated
Z-1023-2014Class IITerminatedVoluntary: Firm initiated
Z-1024-2014Class IITerminatedVoluntary: Firm initiated
Z-1025-2014Class IITerminatedVoluntary: Firm initiated
Z-1026-2014Class IITerminatedVoluntary: Firm initiated
Z-1027-2014Class IITerminatedVoluntary: Firm initiated
Z-1028-2014Class IITerminatedVoluntary: Firm initiated
Z-1029-2014Class IITerminatedVoluntary: Firm initiated
Z-1030-2014Class IITerminatedVoluntary: Firm initiated
Z-1031-2014Class IITerminatedVoluntary: Firm initiated
Z-1032-2014Class IITerminatedVoluntary: Firm initiated
Z-1033-2014Class IITerminatedVoluntary: Firm initiated
Z-1034-2014Class IITerminatedVoluntary: Firm initiated
Z-1035-2014Class IITerminatedVoluntary: Firm initiated
Z-1036-2014Class IITerminatedVoluntary: Firm initiated
Z-1037-2014Class IITerminatedVoluntary: Firm initiated
Z-1038-2014Class IITerminatedVoluntary: Firm initiated
Z-1039-2014Class IITerminatedVoluntary: Firm initiated
Z-1040-2014Class IITerminatedVoluntary: Firm initiated
Z-1041-2014Class IITerminatedVoluntary: Firm initiated
Z-1042-2014Class IITerminatedVoluntary: Firm initiated
Z-1043-2014Class IITerminatedVoluntary: Firm initiated
Z-1044-2014Class IITerminatedVoluntary: Firm initiated