Specialty Straight Static Inserter Assembly
Stryker Spine began a recall of Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to... on . FDA classified it as Class II and gives the reason as "Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1086-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker sent Urgent Medical Device Removal Notification letter and to their customers via email on 12/18/2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory, return affected product using enclosed pre-paid mailing label, complete Customer Response Form and fax to Regulatory Compliance at 855-632-9049 or mail it to [email protected]. For questions call 201-760-8287, 201-760-8298 or 201-760-8041.
- Quantity in commerce
- 21 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1086-2014 | Class II | Terminated | Voluntary: Firm initiated |