Synthes Small Notch Titanium Reconstructive Plate
Synthes, Inc. began a recall of Synthes Small Notch Titanium Reconstructive Plate Product Usage: The Synthes Small Notch Titanium Reconstructive Plate set offers fixation for immediate... on . FDA classified it as Class II and gives the reason as "Certain modules containing the Synthes Small Notch Titanium Reconstructive Plate set was distributed by Spine sales consultants although it was contraindicated for certain spine applications". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1174-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Small Notch Titanium Reconstructive Plate Product Usage: The Synthes Small Notch Titanium Reconstructive Plate set offers fixation for immediate stabilization to allow bone fusion. Commercially pure titanium implants are biocompatible. The wide selection of implants accommodates variable anatomy; precisely crafted instruments facilitate implant placement. Implant plate lateral slant of holes allows for angulation of screws. Elongated central holes provide flexibility in screw placement. Wide selection of length's accommodates individual case requirements.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Certain modules containing the Synthes Small Notch Titanium Reconstructive Plate set was distributed by Spine sales consultants although it was contraindicated for certain spine applications.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Synthes sent an Urgent Medical Medical Device Recall letter, dated January 27, 2013, to Users and Sales Consultants. The letter identified the affected product, reason for recall, potential hazard, and actions to be taken. Customers were instructed to complete the verification section. For questions call 610-719-5450 or contact your Synthes Sales Consultant. .
- Quantity in commerce
- 18747
- Distribution
- USA Nationwide Distribution
- Product codes and lots
- All lots of Parts 489.401, 489.402, 489.412, 489.413, 489.414, 489.415, 489.418, 489.423, 489.424, and 489.425.