510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1174-2014Class II

Synthes Small Notch Titanium Reconstructive Plate

Synthes GmbH / initiated 2014-01-27 / Terminated / root cause: labeling and UDI

Synthes, Inc. began a recall of Synthes Small Notch Titanium Reconstructive Plate Product Usage: The Synthes Small Notch Titanium Reconstructive Plate set offers fixation for immediate... on . FDA classified it as Class II and gives the reason as "Certain modules containing the Synthes Small Notch Titanium Reconstructive Plate set was distributed by Spine sales consultants although it was contraindicated for certain spine applications". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1174-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Small Notch Titanium Reconstructive Plate Product Usage: The Synthes Small Notch Titanium Reconstructive Plate set offers fixation for immediate stabilization to allow bone fusion. Commercially pure titanium implants are biocompatible. The wide selection of implants accommodates variable anatomy; precisely crafted instruments facilitate implant placement. Implant plate lateral slant of holes allows for angulation of screws. Elongated central holes provide flexibility in screw placement. Wide selection of length's accommodates individual case requirements.
Product code
HRS Plate, Fixation, Bone
Reason for recall
Certain modules containing the Synthes Small Notch Titanium Reconstructive Plate set was distributed by Spine sales consultants although it was contraindicated for certain spine applications.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Synthes sent an Urgent Medical Medical Device Recall letter, dated January 27, 2013, to Users and Sales Consultants. The letter identified the affected product, reason for recall, potential hazard, and actions to be taken. Customers were instructed to complete the verification section. For questions call 610-719-5450 or contact your Synthes Sales Consultant. .
Quantity in commerce
18747
Distribution
USA Nationwide Distribution
Product codes and lots
All lots of Parts 489.401, 489.402, 489.412, 489.413, 489.414, 489.415, 489.418, 489.423, 489.424, and 489.425.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1174-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1174-2014Class IITerminatedVoluntary: Firm initiated