Z-0996-2014Class II

Medtronic Select 3D Arterial Cannulae. Vented, model 78622 (22 Fr.). Non-Vented Cannulae, model 78722 (22...

Medtronic, Inc. / initiated 2014-01-30 / Terminated

Medtronic Inc. Cardiac Rhythm Disease Management began a recall of Medtronic Select 3D Arterial Cannulae. Vented, model 78622 (22 Fr.). Non-Vented Cannulae, model 78722 (22 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic... on . FDA classified it as Class II and gives the reason as "Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0996-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic Select 3D Arterial Cannulae. Vented, model 78622 (22 Fr.). Non-Vented Cannulae, model 78722 (22 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic. The Cannula consists of a curved, angled or beveled tip with or without flange that is permanently attached to a molded, clear, flexible PVC tapered body or wirebound PVC body. The cannula body features a tip orientation line to indicate direction of the cannula tip during cannulation. The proximal end of the cannula includes a connection site or a molded 0.95 cm (3/8 in) barbed connector with or without a molded plastic vent plug or peel cap. The vent plug or peel cap allows air to be vented from the cannula before connection to the perfusion line. These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar.
FDA root cause
Under Investigation by firm
Action
The firm, Medtronic, sent an "Urgent Medical Device Recall" letter dated January, 2014 via UPS on January 30, 2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify and quarantine all listed product in your inventory; return all listed product to Medtronic and contact Customer Service at 800 854-3570 and reference RADAR #167826 to initiate a return and credit of product; complete and return the attached Customer Confirmation Certificate via fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality, and share this notification with others in your organization as appropriate. Select 3D and Select CAP Arterial Cannula are not available for purchase at this time. Please contact your Medtronic sales representative or Medtronic LifeLine Technical Services at 1 877-526-7890 or (+1 763) 526-7890 to assist with alternative cannula selection and any additional questions you may have related to this notification.
Quantity in commerce
19483 total (7669 US, 11814 OUS)
Distribution
Worldwide distribution: US (Nationwide) including Puerto Rico, and Internationally to: Algeria, Argentina, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Cyprus, El Salvador, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Jordan, Kuwait, Netherlands, Oman, Paraguay, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Venezuela, and Yemen.
Product codes and lots
Model 78622, Lot numbers: 2011031714, 2011040345, 2011050373, 2011060471,  2011060641, 2011060642, 2011070099, 2011072408, 2011081502, 2011102965,  2011111215, 2011120575, 2012011143, 2012011425, 2012020052, 2012040536,  2012051387, 2012090216, 2012101333 ,2013010957, 2013011791, 2013020303,  2013030729, 2013040209, 2013040282, 2013040710, 2013041272, 2013051447,  2013060736, 2013071836, 2013072009, 201105C105, 201203C854.   Model 78722, Lot numbers: 2011031588, 2011031885, 2011051006, 2011051226,  2011061585, 2011071672, 2011082216, 2011091825 , 2011101670, 2011102093,  2011110717, 2011120059, 2012060478, 2012061056, 2012061057, 2012061725,  2012070570, 2012090657, 2013011121, 2013021260, 2013030685, 2013040494,  2013040595, 2013070162, 2013070483, 2013081289, 2013081603, 201108C227.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0996-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0995-2014Class IITerminatedVoluntary: Firm initiated
Z-0996-2014Class IITerminatedVoluntary: Firm initiated
Z-0997-2014Class IITerminatedVoluntary: Firm initiated
Z-0998-2014Class IITerminatedVoluntary: Firm initiated
Z-0999-2014Class IITerminatedVoluntary: Firm initiated
Z-1000-2014Class IITerminatedVoluntary: Firm initiated