Integra Cusa Excel Rx Only CEM" Nosecone for CUSA Excel System
Integra LifeSciences Corp. began a recall of Integra Cusa Excel Rx Only CEM" Nosecone for CUSA Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System... on . FDA classified it as Class II and gives the reason as "The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0994-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra Cusa Excel Rx Only CEM" Nosecone for CUSA Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.
- Reason for recall
- The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Integra sent an Urgent: Medical Device Correction Letter/Acknowledgement Return Form dated December 5, 2013 via traceable courier to their customers. The notification: advises customers of the nature of the issue and how it could affect CEM Nosecone use, advises customers what to do if the issue occurs and requests customers complete and return the Acknowledgment Return Form. The form should be faxed to [email protected] or 1-609-275-9445. Any questions can be directed to customer service at 855-532-1723.
- Quantity in commerce
- 17253 boxes
- Distribution
- Worldwide Distribution - USA Nationwide
- Product codes and lots
- All unexpired lot numbers for CEM Nosecone Gatalouge No. C6623 and C6636 from 1/1/2010 to present. CEM" Nosecone, Catalogue No. C6623: 1083655, 1084219, 1084398, 1084399, 1084400, 1084682, 1084683, 1090937, 1091712, 1091714, 1091715, 1093125, 1093126, 1093128, 1093129, 1093471, 1093847, 1094575, 1094740, 1095587, 1095873, 1100237, 1100238, 1101217, 1101547, 1101548, 1102069, 1102070, 1102544, 1102545, 1102546, 1103102, 1103392, 1103444, 1103445, 1103446, 1104264, 1104265, 1105065, 1105066, 1105067, 1110208, 1110311, 1110348, 1110349, 1110473, 1110474, 1110892, 1111760, 1112182, 1112808, 1112809, 1113175, 1113176, 1113177, 1113178, 1113662, 1114112, 1114654, 1114655, 1114656, 1115224, 1115225, 1115499, 1115500, 1115501, 1115853, 1115854, 1120771, 1120772, 1120773, 1120774, 1121141, 1121142, 1121383, 1121627, 1121628, 1122257, 1122259, 1122260, 1123050, 1123312, 1123915, 1124642, 1124643, 1124980, 1125393, 1125394, 1125933, 1130036, 1130037, 1130038, 1130039, 1130218, 1130219, 1130663, 1130664, 1131340, 1131344, 1131986, 1132207, 1132326, 1132327, 1132328, 1132880 CEM" Nosecone, Catalogue No. C6636: 1084216, 1084293, 1084377, 1084680, 1084681, 1090123, 1090124, 1090125, 1090126, 1090660, 1090661, 1090826, 1091395, 1091434, 1091700, 1092388, 1092389, 1092390, 1092391, 1092392, 1093012, 1093013, 1093130, 1093131, 1093472, 1093473, 1093474, 1093824, 1093825, 1093826, 1093827, 1094741, 1095104, 1095105, 1095106, 1095414, 1095415, 1095418, 1095419, 1100239, 1100240, 1100241, 1101220, 1101221, 1102647, 1102648, 1102649, 1102650, 1102651, 1103103, 1103394, 1103395, 1103437, 1103848, 1103851, 1103916, 1103917, 1103919, 1103920, 1103921, 1103922, 1104258, 1104259, 1104260, 1104261, 1104262, 1105068, 1105069, 1105070, 1110333, 1110975, 1110976, 1110977, 1111491, 1111492, 1111493, 1112186, 1112187, 1112211, 1112212, 1112769, 1112770, 1112771, 1113180, 1113181, 1113182, 1113183, 1113184, 1113664, 1113665, 1113666, 1114358, 1114359, 1114657, 1114658, 1114659, 1114660, 1115036, 1115037, 1115038, 1115039, 1115503, 1115504, 1115505, 1115855, 1115856, 1115857, 1115858, 1115859, 1115860, 1120360, 1120361, 1120362, 1120775, 1120776, 1120777, 1121568, 1121729, 1122344, 1122346, 1122347, 1123053, 1123054, 1123497, 1123498, 1123499, 1123500, 1123501, 1124054, 1124055, 1124703, 1124704, 1124705, 1124706, 1124707, 1124710, 1125395, 1125396, 1125397, 1125398, 1125399, 1125400, 1125476, 1125934, 1125935, 1125936, 1125937, 1125938, 1125939, 1130220, 1130221, 1130222, 1130223, 1130834, 1131243, 1131987, 1132777, 1132778, 1132779, 1132780, 1132879, 1133061, 1103850, 1103852, 1103917, 1103918, 1103922, 1104257, 1104263, 1105071, 1110475, 1110476, 1110978, 1111490, 1111493, 1112187, 1112188, 1112212, 1113179, 1113180, 1113663, 1114357, 1114358, 1114659, 1114660, 1115039, 1115502, 1115504, 1115505, 1120776, 1120777, 1121167, 1121168, 1121169, 1121565, 1121566, 1121567, 1122344, 1122743, 1123051, 1123052, 1123497, 1123498, 1123499, 1123500, 1123501, 1123508, 1123509, 1123510, 1123511, 1123512, 1124052, 1124053, 1124707, 1124709
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0994-2014 | Class II | Terminated | Voluntary: Firm initiated |