Z-1471-2014Class I

Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device) Catalog Numbers: 1100, 1101, 1102...

HeartWare, Inc. / initiated 2013-12-06 / Terminated / root cause: manufacturing process

HeartWare Inc began a recall of Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device) Catalog Numbers: 1100, 1101, 1102, 1103, 1104, and 1205. For use as a bridge-to-cardiac... on . FDA classified it as Class I and gives the reason as "The housing to the Pump's driveline connector became partially or fully separated from the front portion of the driveline connector". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1471-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
HeartWare, Inc.
Product
Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device) Catalog Numbers: 1100, 1101, 1102, 1103, 1104, and 1205. For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare VAS is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
The housing to the Pump's driveline connector became partially or fully separated from the front portion of the driveline connector.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
HeartWare, Inc. sent an Urgent Medical Device Correction letter to all their hospitals and doctors on December 12, 2013 instructing that at each routine clinic visit, they should inspect the patient's driveline connector as described in the attached Technical Bulletin. For already implanted patients, arrange a follow up visit at the earliest convenience to check patient's driveline connector. They were asked to sign and return the Acknowledgement Form to HeartWare. For questions, contact your HeartWare Representative or HeartWare Customer Service at 1-877-367-4823.
Quantity in commerce
4,586
Distribution
Worldwide Distribution - USA Nationwide in the states of MN, IL, FL, WI, MO, MA, NC, CA, GA, AL, CO, TX, OH, NY, VA, UT, KY, MD, AZ, SC, IN, PA, DE, NJ, LA, OR, WA, MI, IA, TN and DC and countries: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Singapore, Slovakia, South Africa, Spain, Switzerland, Turkey and United Kingdom.
Product codes and lots
Catalog Numbers: 1100, 1101, 1102, 1103, 1104, and 1205. Serial Numbers: HW001 to HW11270 and HW20001 to HW 20296.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1471-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1471-2014Class ITerminatedVoluntary: Firm initiated