FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is...
Djo Surgical began a recall of FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is used with an acetabular shell to replicate the... on . FDA classified it as Class II and gives the reason as "Packaging error -- two different types and sizes of acetabular liners used during hip replacement surgery were mistakenly packaged into each other's outer boxes after sterilization, resulting in the products being...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1130-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is used with an acetabular shell to replicate the kinematics of a natural hip.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Packaging error -- two different types and sizes of acetabular liners used during hip replacement surgery were mistakenly packaged into each other's outer boxes after sterilization, resulting in the products being mislabeled. The units labeled as Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 on the outside and containing Model/Catalog Number 932-36-252, Lot# 728F1089 on the inside (1
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- DJO Global sent an Urgent Field Safety Notice dated February 7, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. All 10 devices with the outer label of 931-28-248, Lot#685F1034 were found in-house and were quarantined. The agencies who have the affected product (devices labeled as 932-36-252, Lot# 728F1089 containing 931-28-248, Lot#685F1034) were contacted to quarantine their devices. Of the affected devices, 2 of the lot of 10 have already been returned via product complaint. Customers were instructed to pass on the notice to all those who need to be aware within their organization or to any organization where the potentially affected devices have been transferred. Customers were instructed to place a replacement order and receive an RMA number to return all affected devices.. Customers with questions were instructed to call 512-834-6255.
- Quantity in commerce
- 10 units of Model/Catalog Number 931-28-248 Lot/Serial Number(s) 685F1034 labeled asModel/Catalog Number 932-36-252, Lot# 728F1089
- Distribution
- Nationwide Distribution - USA including TX, LA, FL, CA, and AZ.
- Product codes and lots
- Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 (an MP5 Non-Hooded, Neutral Liner) were mistakenly packaged/labeled as Model/Catalog Number 932-36-252, Lot# 728F1089 (an MP7 10-degree Hooded Liner) and vice-versa
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1130-2014 | Class II | Terminated | Voluntary: Firm initiated |