Z-1155-2014Class II

Philips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A...

Philips North America LLC / initiated 2014-02-18 / Terminated / root cause: design

Philips Medical Systems, Inc. began a recall of Philips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A, M3536M,M3536M4,M3536M6,861288,861289,861464,861483,and 861491 on . FDA classified it as Class II and gives the reason as "MRx may display a Red X in the Ready For Use (RFU) indicator during automated tests informing the user that a failure has been detected that may prevent the delivery of a shock and pacing". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1155-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A, M3536M,M3536M4,M3536M6,861288,861289,861464,861483,and 861491
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
MRx may display a Red X in the Ready For Use (RFU) indicator during automated tests informing the user that a failure has been detected that may prevent the delivery of a shock and pacing
Root cause tag
design
FDA root cause
Component design/selection
Action
Philips Healthcare issued and Urgent Medical Device Correction notification/Field Safety Notice on 02/18/14 to inform customers of the issue, Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: You can continue to use your MRx prior to receiving the hardware upgrade, provided that if you observe that the MRx is displaying a Red X in the Ready For Use (RFU) indicator\during automated tests, a failure has been detected that may prevent the delivery of a shock or pacing due to this issue. If this is observed you should turn the Therapy Knob to Monitor. An INOP message describing the failure is displayed. If needed, run an Operational Check for further information. If the condition persists, take the device out of use and call for service. Identify a readily available backup defibrillator to use in the event the affected HeartStart MRx fails to deliver therapy. Philips will perform a hardware upgrade (replace therapy boards) free of charge to be installed by a Philips Field Service Engineer. Contact your local Philips representative or call us at 1-800-722-9377.
Quantity in commerce
3541 units
Distribution
Nationwide Foreign: AUSTRIA BAHRAIN BELGIUM BRAZIL CANADA CHINA EGYPT ESTONIA FINLAND FRANCE GERMANY GHANA ICELAND INDIA INDONESIA IRELAND ISRAEL ITALY JAPAN JORDAN KENYA KOREA, REPUBLIC OF LATVIA LEBANON LIBYAN ARAB JAMAHIRIYA LITHUANIA MALAYSIA MEXICO MOZAMBIQUE MYANMAR (Burma) NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NORWAY PAKISTAN PERU POLAND PORTUGAL QATAR RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN, PROVINCE OF CHINA THAILAND TURKMENISTAN UNITED ARAB EMIRATES UNITED KINGDOM VIET NAM
Product codes and lots
Serial numbers within the range: US00100204 to US00570921

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1155-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1155-2014Class IITerminatedVoluntary: Firm initiated