510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1157-2014Class II

Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction...

Smith & Nephew, Inc. / initiated 2014-01-09 / Terminated / root cause: packaging

Smith & Nephew Inc. began a recall of Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps) on . FDA classified it as Class II and gives the reason as "Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1157-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
Product code
OMP Negative Pressure Wound Therapy Powered Suction Pump
Reason for recall
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
Root cause tag
packaging
FDA root cause
Packaging change control
Action
A product bulletin titled "Soft Port Enhancements" dated March 14, 2013, was released for external distribution describing the changes to the Soft Port products, e.g. increasing the Soft Port orifices size and adding a second slit to the release paper covering the port adhesive, and stating the purpose of the modifications to the device. These product bulletin was provided to sales representatives for distribution to customers requesting information regarding the design enhancement made to the Soft Port product range. The bulletin was not distributed to all the customers. To ensure that the product modification and its purpose was appropriately communicated to consignees of affected product a Medical Device Correction Notice, dated 01/22/2014, retrospectively advising customers of the modification to the Soft Port products has been prepared and was sent to consignees of the affected products on the same date.
Quantity in commerce
162,629 invididual units
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Product codes and lots
ALL LOT #s 2013010311 and lower

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1157-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1157-2014Class IITerminatedVoluntary: Firm initiated
Z-1158-2014Class IITerminatedVoluntary: Firm initiated
Z-1159-2014Class IITerminatedVoluntary: Firm initiated
Z-1160-2014Class IITerminatedVoluntary: Firm initiated
Z-1161-2014Class IITerminatedVoluntary: Firm initiated
Z-1162-2014Class IITerminatedVoluntary: Firm initiated