Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port...
Smith & Nephew Inc. began a recall of Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative... on . FDA classified it as Class II and gives the reason as "Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1162-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
- Product code
- OMP Negative Pressure Wound Therapy Powered Suction Pump
- Reason for recall
- Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- A product bulletin titled "Soft Port Enhancements" dated March 14, 2013, was released for external distribution describing the changes to the Soft Port products, e.g. increasing the Soft Port orifices size and adding a second slit to the release paper covering the port adhesive, and stating the purpose of the modifications to the device. These product bulletin was provided to sales representatives for distribution to customers requesting information regarding the design enhancement made to the Soft Port product range. The bulletin was not distributed to all the customers. To ensure that the product modification and its purpose was appropriately communicated to consignees of affected product a Medical Device Correction Notice, dated 01/22/2014, retrospectively advising customers of the modification to the Soft Port products has been prepared and was sent to consignees of the affected products on the same date.
- Quantity in commerce
- 28,250 individual units
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
- Product codes and lots
- ALL LOT #'s 2013010287 and lower
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1157-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1158-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1159-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1160-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1161-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1162-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.