Z-1108-2014Class II

M3536A HeartStart MRx with Wireless Link Upgrade (861289) The HeartStart MRx is for use for the termination...

Philips North America LLC / initiated 2014-02-18 / Terminated / root cause: software

Philips Medical Systems, Inc. began a recall of M3536A HeartStart MRx with Wireless Link Upgrade (861289) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular... on . FDA classified it as Class II and gives the reason as "Wireless Link may be unable to transmit data when configured for a specific data flow". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1108-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
M3536A HeartStart MRx with Wireless Link Upgrade (861289) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Wireless Link may be unable to transmit data when configured for a specific data flow.
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare sent an Urgent Medical Device Correction letter to all consignees on February 18, 2014. The Letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips is initiating a correction to affected devices. Philips will release a new version of Wireless Link software V2.0.0.3 sub1 that corrects the issue. The correction will be provided free of charge to all units affected by this issue. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the software upgrade. Customers with questons were instructed to call 1-800-722-9377. For questions regarding this recall call 800-722-9377.
Quantity in commerce
220 devices
Distribution
Worldwide Distribution - USA including AZ, CA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX,and VA and Internationally to Australia, Norway, and Spain.
Product codes and lots
US00100253, US00547494, US00565643, US00567331, US00100950, US00558361, US00565651, US00567332, US00204282, US00560173, US00565652, US00567333, US00210992, US00560174,US00565653, US00567334, US00210993, US00560175, US00565655, US00567335, US00210994, US00560176, US00565656, US00567336, US00210995, US00560177, US00565657, US00567337, US00210996, US00560178, US00565658, US00567338, US00210997, US00561068, US00565660, US00567339, US00210998, US00561724, US00565661, US00567340,  US00210999, US00563602, US00565662, US00567341, US00211000, US00563603, US00565663, US00567720, US00211001, US00563606, US00565664, US00567721, US00212264, US00563607, US00565665, US00567722, US00212265, US00565069, US00565666, US00567723,  US00212266, US00565414, US00566434, US00567724, US00212267, US00565415, US00566564, US00568419, US00212268, US00565416, US00567302, US00568656, US00322574, US00565417, US00567303, US00568703, US00322575, US00565418, US00567304, US00568704 US00322876, US00565419, US00567305, US00568705, US00322877, US00565420, US00567306, US00568706, US00323134, US00565421, US00567307, US00568748, US00323135, US00565422, US00567308, US00568749, US00323137, US00565423, US00567309, US00568750,  US00323138, US00565424, US00567310, US00568751, US00541205, US00565425, US00567311, US00568752, US00541250, US00565427, US00567312, US00568753, US00541254, US00565428, US00567313, US00568754, US00541364, US00565429, US00567314, US00568755,  US00541370, US00565430, US00567315, US00568756, US00541397, US00565446, US00567316, US00568757, US00541402, US00565447, US00567317, US00568758, US00541403, US00565448, US00567318, US00568759, US00541405, US00565449, US00567319, US00568760,  US00543683, US00565450, US00567320, US00568761, US00543684, US00565451, US00567321, US00568762, US00543860, US00565452, US00567322, US00568763, US00544078, US00565453, US00567323, US00568764, US00544082, US00565454, US00567324, US00568765,  US00544089, US00565455, US00567325, US00568766, US00544095, US00565456, US00567326, US00568767, US00547383, US00565457, US00567327, US00568768, US00547456, US00565458, US00567328, US00568769, US00547457, US00565459, US00567329, US00568770, US00547458, US00565642, US00567330, US00568771

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1108-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1108-2014Class IITerminatedVoluntary: Firm initiated