Z-1109-2014Class II

Philips HeartStart XL, Defibrillator/Monitor Model number M4735A

Philips North America LLC / initiated 2014-02-14 / Terminated / root cause: design

Philips Medical Systems, Inc. began a recall of Philips HeartStart XL, Defibrillator/Monitor Model number M4735A Product Usage: The M4735A HeartStart XL Defibrillator/Monitor is for use in the hospital by... on . FDA classified it as Class II and gives the reason as "Philips HeartStart XL may experience a power board failure causing failure to defibrillate". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1109-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips HeartStart XL, Defibrillator/Monitor Model number M4735A Product Usage: The M4735A HeartStart XL Defibrillator/Monitor is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician. When operating as a semi-automatic external defibrillator in AED Mode, the M4735A HeartStart XL Defibrillator/Monitor is suitable for use by medical personnel trained in basic life support that includes the use of an AED
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Philips HeartStart XL may experience a power board failure causing failure to defibrillate
Root cause tag
design
FDA root cause
Component design/selection
Action
Philips Healthcare issued an Urgent Medical Device Correction notification/Field Safety Notice dated February 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed Philips Healthcare Representative will perform a hardware upgrade free of charge. Customers were instructed while awaiting the replacement power board for your device, you may continue to use the HeartStart XL. Identify a readily available backup defibrillator to use in the event the affected HeartStart XL fails to deliver therapy. If you need any further information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377
Quantity in commerce
5234 units
Distribution
Worldwide Distribution and USA Nationwide and the countries of AUSTRALIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, ETHIOPIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, INDIA, INDONESIA, IRELAND, ITALY, JAPAN, KAZAKHSTAN, KOREA, REPUBLIC OF, MALAYSIA, MEXICO, MONACO, NEPAL, NETHERLANDS, PORTUGAL, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, UNITED KINGDOM, UZBEKISTAN, VENEZUELA, and VIET NAM,
Product codes and lots
Serial numbers within the range:  US00101665 through US00608704 Units manufactured by Philips between May, 2013 and November, 2013.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1109-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1109-2014Class IITerminatedVoluntary: Firm initiated