Z-1156-2014Class II

Panaroma Central Station

Datascope Corp. / initiated 2014-01-08 / Terminated / root cause: software

Mindray DS USA, Inc. dba Mindray North America began a recall of Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can... on . FDA classified it as Class II and gives the reason as "Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindrays V-Series Monitor...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1156-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindrays V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.
Root cause tag
software
FDA root cause
Software design
Action
Mindray DS USA Inc. sent an Urgent Corrective Action letter dated January 8, 2014 via certified mail to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to contact Mindray Service Representative, at 1-800-288-2121 (Monday Friday 8:30 am-5:30 pm), to arrange for the software update of the Panorama System(s) in their facility. This work will be performed at no cost.
Quantity in commerce
732 units in the US
Distribution
Worldwide Distribution - USA Nationwide Distribution
Product codes and lots
p/n 0998-00-0708-01; 0998-UC-0708-01 and 0020-00-0205-XXX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1156-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1156-2014Class IITerminatedVoluntary: Firm initiated