510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1251-2014Class II

Natural-Knee II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONG...

Zimmer GmbH / initiated 2014-02-25 / Terminated / root cause: packaging

Zimmer, Inc. began a recall of Natural-Knee II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONG TIB INS SZ1/2 13MM U CONG TIB INS SZ3/4 9MM UN... on . FDA classified it as Class II and gives the reason as "A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1251-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Natural-Knee II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONG TIB INS SZ1/2 13MM U CONG TIB INS SZ3/4 9MM UN CONG TIB INS SZ3/4 11MM U CONG TIB INS SZ3/4 13MM U CONG TIB INS SZ 5/6 9MM U CONG TIB INS SZ5/6 11MM U CONG TIB INS SZ5/6 13MM U
Product code
HRY Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer
Reason for recall
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Root cause tag
packaging
FDA root cause
Storage
Action
Beginning 2/25/2014 all distributors were notified via electronic mail. Hospital risk managers and distributors, surgeons with affected inventory will also be notified via courier or direct mail. The Urgent Lot Specific Recall Notice identified the issue and responsibilities. Distributors were asked to locate and remove the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product. Distributors were asked to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter and email it back to [email protected]. Hospital risk managers and surgeons were asked to locate affected product and return it to their Zimmer sales representative. Questions or concerns should be addressed to customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST Communications outside of the United States will occur approximately two weeks after the United States communications. 100 % effective checks will be done.
Quantity in commerce
3638 units
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Product codes and lots
Item Number 611001009 611001011 611001013 611002009 611002011 611002013 611003009 611003011 611003013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1251-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1248-2014Class IITerminatedVoluntary: Firm initiated
Z-1249-2014Class IITerminatedVoluntary: Firm initiated
Z-1250-2014Class IITerminatedVoluntary: Firm initiated
Z-1251-2014Class IITerminatedVoluntary: Firm initiated
Z-1252-2014Class IITerminatedVoluntary: Firm initiated
Z-1253-2014Class IITerminatedVoluntary: Firm initiated
Z-1254-2014Class IITerminatedVoluntary: Firm initiated
Z-1255-2014Class IITerminatedVoluntary: Firm initiated
Z-1256-2014Class IITerminatedVoluntary: Firm initiated
Z-1257-2014Class IITerminatedVoluntary: Firm initiated
Z-1258-2014Class IITerminatedVoluntary: Firm initiated
Z-1259-2014Class IITerminatedVoluntary: Firm initiated
Z-1260-2014Class IITerminatedVoluntary: Firm initiated
Z-1261-2014Class IITerminatedVoluntary: Firm initiated
Z-1262-2014Class IITerminatedVoluntary: Firm initiated