Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1260-2014Class II

Apollo" Knee System: PS TIB INS SZ0/16MM APOLL PS TIB INS SZ0/19MM APOLL PS TIB INS SZ0/22MM APOLL PS TIB INS...

Zimmer GmbH / initiated 2014-02-25 / Terminated / root cause: packaging

Zimmer, Inc. began a recall of Apollo" Knee System: PS TIB INS SZ0/16MM APOLL PS TIB INS SZ0/19MM APOLL PS TIB INS SZ0/22MM APOLL PS TIB INS SZ1/16MM APOLL PS TIB INS SZ2/16M APOLLO PS TIB... on . FDA classified it as Class II and gives the reason as "A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1260-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Apollo" Knee System: PS TIB INS SZ0/16MM APOLL PS TIB INS SZ0/19MM APOLL PS TIB INS SZ0/22MM APOLL PS TIB INS SZ1/16MM APOLL PS TIB INS SZ2/16M APOLLO PS TIB INS SZ2/22MM APOLL PS TIB INS SZ3/16MM APOLL PS TIB INS SZ3/19MM APOLL PS TIB INS SZ4/9MM APOLLO PS TIB INS SZ4/11MM APOLL PS TIB INS SZ4/13MM APOLL PS TIB INS SZ4/16MM APOLL PS TIB INS SZ4/19MM APOLL PS TIB INS SZ4/22MM APOLL PS TIB INS SZ5/9MM APOLLO PS TIB INS SZ5/19MM APOLL
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Root cause tag
packaging
FDA root cause
Storage
Action
Beginning 2/25/2014 all distributors were notified via electronic mail. Hospital risk managers and distributors, surgeons with affected inventory will also be notified via courier or direct mail. The Urgent Lot Specific Recall Notice identified the issue and responsibilities. Distributors were asked to locate and remove the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product. Distributors were asked to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter and email it back to [email protected]. Hospital risk managers and surgeons were asked to locate affected product and return it to their Zimmer sales representative. Questions or concerns should be addressed to customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST Communications outside of the United States will occur approximately two weeks after the United States communications. 100 % effective checks will be done.
Quantity in commerce
3638 units
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Product codes and lots
Item Number Item Number 681600016 681600019 681600022 681601016 681602016 681602022 681603016 681603019 681604009 681604011 681604013 681604016 681604019 681604022 681605009 681605019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1260-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1248-2014Class IITerminatedVoluntary: Firm initiated
Z-1249-2014Class IITerminatedVoluntary: Firm initiated
Z-1250-2014Class IITerminatedVoluntary: Firm initiated
Z-1251-2014Class IITerminatedVoluntary: Firm initiated
Z-1252-2014Class IITerminatedVoluntary: Firm initiated
Z-1253-2014Class IITerminatedVoluntary: Firm initiated
Z-1254-2014Class IITerminatedVoluntary: Firm initiated
Z-1255-2014Class IITerminatedVoluntary: Firm initiated
Z-1256-2014Class IITerminatedVoluntary: Firm initiated
Z-1257-2014Class IITerminatedVoluntary: Firm initiated
Z-1258-2014Class IITerminatedVoluntary: Firm initiated
Z-1259-2014Class IITerminatedVoluntary: Firm initiated
Z-1260-2014Class IITerminatedVoluntary: Firm initiated
Z-1261-2014Class IITerminatedVoluntary: Firm initiated
Z-1262-2014Class IITerminatedVoluntary: Firm initiated