Z-1308-2014Class II
RaySearch Ray Station, Radiation Therapy Treatment Planning System, Stand-alone software
RaySearch Laboratories AB (publ) / initiated 2014-02-06 / Terminated / root cause: software
RAYSEARCH LABORATORIES AB began a recall of RaySearch Ray Station, Radiation Therapy Treatment Planning System, Stand-alone software on . FDA classified it as Class II and gives the reason as "RaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1308-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RaySearch Ray Station, Radiation Therapy Treatment Planning System, Stand-alone software
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- RaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- On February 6, 2014, RaySearch Laboratories distributed Urgent Medical Device Correction notices to their customers via email regarding the affected product. To determine if the version you are using is affected, bring up the About box in the RayStation application and check if the number reported is "4.0.2.9". If so, the notice applies to your version. Customers are advised to inform all planning staff and all users about the instructions listed on the notice. Customers are advised to inspect their system and identify all installed units with the referenced software version numbers listed on the notice. Corrective labeling is distributed to the customers in the form of the safety notification attached to this report and the software problem will be resolved in RayStation patch 4.0.3 scheduled for market release end of February 2014 and in all future releases. For further information and return of the reply form, please contact: Freddie Cardel, Director of Customer Support, RaySearch Americas, (877) 778-3849 or [email protected].
- Quantity in commerce
- 19 units (16 units domestically & 3 units internationally)
- Distribution
- OH, MD, NJ, WA, IL, MA, FL, Spain, United Kingdom, Italy,
- Product codes and lots
- Model#: 4.30; Version: 4.0.2.9; 3D-CRT optimization or Fallback Planning Licenses
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1308-2014 | Class II | Terminated | Voluntary: Firm initiated |