Z-1306-2014Class II

i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of...

Abbott Medical / initiated 2014-01-01 / Terminated / root cause: manufacturing process

Abbott Point Of Care Inc. began a recall of i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and... on . FDA classified it as Class II and gives the reason as "Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1306-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
APOC sent an Urgent Recall Notice dated January 2014 via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product and provide a copy of this letter to any institution that was forwarded the affected product. All unused cartridges will be credited and are to be returned as instructed on the enclosed Business Reply Card. To ensure continuity of supply at their facility, customers were instructed to follow their standard process to place an order or call Customer Sedrvice at 1-800-323-9100, Option #2, then Option #9). Customers with questions were instructed to contact Abbott Point of Care Technical Support at 1-800-366-8020, Option 1 of their Abbott Point of Care representative. For questions regarding this recall call 800-323-9100.
Quantity in commerce
48, 450 US, 1400 OUS (49,850 catridges)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Product codes and lots
List number 03P78-25 :Lot number N13183

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1306-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1306-2014Class IITerminatedVoluntary: Firm initiated