AXIOM Artis Systems
Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and... on . FDA classified it as Class II and gives the reason as "Certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1272-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens sent a recall notification letter, dated February 7, 2014, to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions call +49 9191 18 0.
- Quantity in commerce
- 37
- Distribution
- USA Nationwide Distribution in the states of WI, KY, OH, NJ, NC, FL, PA, MA, IA, TX, NY, IL, MI, AZ, CA, and LA.
- Product codes and lots
- Model nos. 7412807, 7727717, 7728392 with serial numbers: 35020, 34999,35098 35113 35042 35061 35064 35073 35014 35069 35053 35055 35056 35054 35008 35057 35044 35018 35097 35108 35046 35048 35050 35045 35004 35019 46004 46005 40000 40009 40011 40006 40007 40010 44016 44005 44011
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1272-2014 | Class II | Terminated | Voluntary: Firm initiated |