PowerPort Slim Implantable Port, indicated for patient therapies requiring repeated access to the vascular...
Bard Access Systems began a recall of PowerPort Slim Implantable Port, indicated for patient therapies requiring repeated access to the vascular system. Product Codes: 1716000, 1716070 on . FDA classified it as Class II and gives the reason as "Labeling discrepancy.The label states that the implantable port contains silicone suture plugs, however approximately 23% of the units contained in the affected lots have a PowerPort Slim titanium implantable port...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1300-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- PowerPort Slim Implantable Port, indicated for patient therapies requiring repeated access to the vascular system. Product Codes: 1716000, 1716070
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- Labeling discrepancy.The label states that the implantable port contains silicone suture plugs, however approximately 23% of the units contained in the affected lots have a PowerPort Slim titanium implantable port without silicone suture plugs.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Notified by Urgent Customer Notification Letter date February 13, 2014, customers were asked to immediately examine their inventory and identify any product subject to this notification. In addition, if they have further distributed this product, please identify customers and notify them at once of this notification. Customers may include a copy of this letter in your notification to customers. Customers may choose to use the involved product with the understanding that the kit may contain a PowerPort Slim implantable port without suture plugs, or may return the applicable product for replacement or credit.
- Quantity in commerce
- 777
- Distribution
- US Nationwide distribution including WA, KY, IL, and IA.
- Product codes and lots
- Lot Numbers: REXA1302, REXA1350, REXB1160